ESWL (Extracorporal Shockwave Lithotripsy)
Main Functions:
- Shock wave source: Low energy electromagnetic shock wave source.
- Localization: C arm X-ray & B-ultrasound localization.
- Incident angle: Adjustable angle of the shock wave source. Shock
wave source connects to its holder on small C arm, which can be
moved away from the table.
- Table: Seperate and movable table can be used for micro-invasive
operation, scope operation and diagnosis.
- Image: It adopts main frequency & high frequency X-ray system,
high-quality image intensifier, CCD camera and digital
multi-functional monitor.
- Operating system: Computer control and manual control are combined
together, so mouse, keyboard and buttons can be used to operate
the lithotripter. It also contains image workstation. The main
malfunction can be alarmed by the computer.
- Environment: Small size, light, small space needed and 220V working
power supply.
Standard Parameters:
High voltage discharging range |
5~11kV
(Electromagnetic:12~18kV) |
High voltage discharging capacity storage |
7.2~20J
(Electromagnetic:46~104J) |
Pressure parameters on second focus |
|
Pulse front |
≤0.4μS |
Pulse width |
≤0.8μS |
Focal area size |
7*7*25mm |
Height of second focus |
130±2mm |
Movement of the table |
X≤120mm, Y≤120mm, Z≤150mm |
Error of the second focus caused by the movement of the C arm |
≤±2mm |
Maximum load capacity of the table |
135kG |
Standard Configuration:
- Main unit
- Table
- Electrical assembly
- Console
- X-ray assembly
- Image intensifier
- CCD camera
- Monitor
- Shock wave source
- Cushion
- Auxiliaries
[Back]
|